Response; Power Supply; Isolation - Knick SensoTrans R A 20230P0 Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Response

Characteristic
Measuring rate
Response time t 99 *
* Time after change of input value until reaching an output value of 99 % steady
state

Power Supply

Power supply

Isolation

Test voltage
Working voltage
(basic insulation)
Rising / falling linearly
Approx. 3/s
Approx. 2 /s in the operating modes
3-wire resistance measurement,
resistance measurement for R > 5 kΩ
and 4-wire ratio measurement
300 ms
500 ms in the operating modes
3-wire resistance measurement,
resistance measurement for R > 5 kΩ
and 4-wire ratio measurement
24 V DC (-20 %, +25 %)
approx. 0.85 W
2.5 kV, 50 Hz: power supply against
input against output
Up to 300 V AC/DC across all circuits
with overvoltage category II and
pollution degree 2. For applications
with high working voltages, take
measures to prevent accidental
contact and make sure that there
is suffi cient distance or insulation
between adjacent devices.
Specifi cations
49

Publicité

Table des Matières
loading

Table des Matières