Trouble-shooting:
Faults experienced
Solution/how to resolve the problem
Check it has been correctly connected.
The meter isn't
turning on.
Check that the USB port used (computer,
mains adapter...) is receiving power.
Press the ON button.
Proceed as described in step 2.
Visual color alarm does
not display.
Yellow visual alarm must be set up
first.
Alarm threshold are not
Red visual alarm must be set up last.
correct.
Red alarm value must be higher than
yellow alarm value.
Proceed as described in step 2:
The alarm doesn't
sound alarm threshold corresponds to
sound.
red visual alarm value only.
Make sure the sound of the alarm has
not been disabled. Enable the sound
if necessary by pressing the
designated button (step 3).
Check that the USB and micro USB
The battery isn't
ports are properly connected.
charging.
Check that the USB port used
(computer, mains adapter...) is
receiving power.
Imported by:
-57-
AIC International - 44980 Ste Luce sur Loire - France
Faults experienced
Solution/how to resolve the problem
Use the cable provided. The meter
The software does
must be properly switched on (ON
not recognise the
position) to be recognised.
meter.
Try the following two steps:
Turn on the meter and wait for it to
warm up. Connect the USB cable and
start the software.
Turn on the meter and wait for it to
warm up. Start the software. Then
connect the USB cable.
Go to step 6 to set the recording
The measurements
parameters.
aren't being record-
Follow step 7 properly to save the
recording.
ed/saved.
Turn the monitor off for a few minutes,
The meter displays
then on again and place it outside
frozen or erroneous
(without sources of pollution) for 30
values
minutes to clear the sensors.
If values remain incorrect, proceed to
calibration.