Recommended Application Range - VITLAB Dispenser TA2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3. Recommended Application Range

Select valves with platinum-iridium or tantalum valve springs, based on the intended application.
The instrument can be used for the following dispensing media:
Dispensing medium
Acetic acid
Ammonia solution
Bromine
Hydrochloric acid
Hydrofluoric acid*
Hydrogen peroxide
Nitric acid
Perchloric acid
Phosphoric acid
Sodium hydroxide, 30 %
Sulfuric acid
Water
+ suitable
- not suitable
* Note:
Hydrofluoric acid reacts slightly with sapphire resulting in mildly elevated aluminum values. To reduce
these values we recommend discarding 3-5 dispensings of 2 ml each before performing the analysis.
Compounds containing fluorine, such as NaF, attack tantalum.
The above recommendations reflect testing completed prior to publication. Always follow instructions in the
operating manual of the instrument as well as the reagent manufacturer's specifications. In addition to these
chemicals, a variety of organic and inorganic saline solutions (e.g., biological buffers), biological detergents
and media for cell culture can be dispensed. Please call us if you need information on chemicals that are not
named in the list. Status as of: 0216/2
30
Valve springs: Pt-Ir
+
+
+
+
+
-
+
+
+
+
+
+
Valve springs: Ta
+
+
+
+
-
+
+
+
+
-
+
+

Publicité

Chapitres

Table des Matières
loading

Table des Matières