Motor Protection; Error Codes - IKA MTS 2/4 digital Mode D'emploi

Masquer les pouces Voir aussi pour MTS 2/4 digital:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

While Timer mode is in operation, you can increase or decrease the time
setting as required by pressing the "1" key (6) and/or the "10" key (5)
as described above. After you have changed the time setting, the time
begins to elapse again and the LED display (2) shows the remaining time
in minutes as it did before.

Motor protection

Blocking or overloading the motor beyond the permissible motor tem-
perature causes the device to switch off automatically due to the safety
circuit. The LED display (2) displays the corresponding error message. To
eliminate the error, please check the operating conditions. Also allow
the device to cool off.
The device must be turned off and back on again.

Error codes

Error message MTS 2/4:
Please note! If an error occurs, an audio warning signal is generated and
the Timer display switches to E + error number. First try to turn off the
device and then turn it back on to continue operation. If an error is not
eliminated even after waiting for a considerable amount of time, please
contact our service department.
You should always let us know what error code was displayed.
This facilitates troubleshooting and helps us draw some initial conclu-
sions.
Error message
Error
E 1
The potentiometer for the -Too rapida change in the
speed setting does not
specify any target value
E 2
Motor blocked
Read fork signal not
present
E 3
Maximum adjustable
speed of the device
exceeded
E 4
Motor standstill for de-
vice switch-on routine
(safety query)
not provided
E 5
Safety relay cannot be
activated
E 7
Problem with power
supply
E 8
Triac cannot be activated
Cause of error
speed target value
-Internal device error
-Agitation table is inhibi-
ted in its stroke motion by
an external effect
-Internal device error
-Resonance behavior
of the test setup
-To switching on the
device, the motor does
not en rotation motion
-Internal device error
-The device is being ope-
rated with over-voltage or
under-voltage. (For per-
missible voltage range,
see the Technical Data)
-Internal device error
11

Publicité

Table des Matières
loading

Table des Matières