AcceleDent OPTIMA Mode D'emploi page 17

Masquer les pouces Voir aussi pour OPTIMA:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10.4. Transport and Storage Conditions
Ambient Temperature Range: -20 to 50°C
Relative Humidity Range: 30 - 93% RH
10.4.1. If AcceleDent Optima has been transported or stored at its maximum
or minimum transport or storage conditions, the product shall be allowed
to cool or warm for at least two hours within the environmental operating
conditions before use.
10.5. The Activator is a Type BF Applied Part, meaning this medical device
has direct contact with the patient. Such devices shall be marked with
symbol.
10.6. A Power Adapter intended for connection to the USB port of the Charging
Case shall comply with the relevant product standard e.g. IEC 60950-1 or
IEC 62368-1 for IT-equipment and the IEC 60601-series for Medical Electrical
Equipment. In addition, all such combinations – Medical Electrical Systems
– shall comply with the safety requirements stated in the general standard
IEC 60601-1, edition 3, clause 16. Any equipment not complying with the
leakage current requirements in IEC 60601-1 shall be kept outside the patient
environment i.e. at least 1.5 m from the patient support or shall be supplied
via a separation transformer to reduce the leakage currents.
10.7. Use the provided Power Adapter or one that is certified as Class II
adapter with double or reinforced insulation. Such Power Adapter shall be
marked with
symbol or labeled Class II. The specifications of an appropriate
Power Adapter are detailed below:
10.7.1. Certification: Class II, IEC 60950
Output Voltage: 5 V
Output Current Limit: 1 A
AC Input Voltage Range: 100-240 V
AC Input Current: 0.2 A
AC Input Frequency: 50/60 Hz
10.8. Any person who connects external equipment to the USB connector
has formed a Medical Electrical System and is therefore responsible for the
system to comply with the requirements. If in doubt, contact a qualified medical
technician or your local representative.
10.9. The USB port located on the back of the Charging Case is for battery
charging purposes only. The Charging Case has no data communication
BACK TO CONTENTS
17
Page
of 51

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

405-1306-001 d

Table des Matières