Vibration Isolator Installation; Purge Valve Installation; Serial Cable Installation - Varian TV 1001 Navigator Mode D'emploi

Table des Matières

Publicité

VIBRATION ISOLATOR INSTALLATION

Three vibration isolators for ISO and CFF inlet
flange version pumps are available as accesso-
ries.
The three model part numbers are the following:
− model 969-9345 for ISO 160 flange;
− model 969-9346 for ISO 200 flange;
− model 969-9336 for CFF 10" flange.
They typically reduce the vibration transmitted
from the TV 1001 Navigator to the system by a
factor of 20.
Please refer to the relevant instruction manual.

PURGE VALVE INSTALLATION

A gas purge valve is available to protect the pump
bearings against particulate and corrosive gases
that could move into the pump.
To install the gas purge valve it is necessary to
unscrew the purge port cover as shown in the
following figure,
and then screw the gas purge valve (with a torque
of 2.5 Nm) as shown in the following figure.

SERIAL CABLE INSTALLATION

The supplied serial cable must be installed when
the TV 1001 Navigator or the optional vent valve
have to be controlled by means of a remote com-
puter.
The cable is installed fixing the 9 pin D-type con-
nector into the P2 serial connector as shown in the
following figure.
87
TECHNICAL INFORMATION
87-900-945-01(G)

Publicité

Table des Matières
loading

Table des Matières