Conductivity
Temperature
Resolution
Inputs
Temp. Comp.
Storage memory
Analogue output raw electrode signal
Display
Keys
Ambient temp.
Rel. Humidity
Power supply
Cabinet
Dimensions
Weight
Each time the instrument is switched on, it will check briefly its internal
calibrations while showing all functions of the display and the software-
version. An error message appears when sonething is wrong.
MODE
= Selects all modes or escapes from error traps, calibration
procedures, etc..by returning to the original mode.
CAL
= Starts or proceeds a calibration or a function.
/
= Button for entering a value or for selecting a function.
ON/OFF
= Switches the instrument on or off.
[Or]
= Overrange (e.g. input short circuited).
[Err.cc]
= Cell constant not within 0.7...1.3 cm
[CAL]
= Calibration error (check electrodes and standard solutions).
[Err.MEM] = Fatal memory error (all factory settings are cleared. Have
your instrument serviced!).
0...100 mS/cm
0...100°C
0.5°C, 0.1 µS/cm
BNC input for conductivity cell
BANANA input for a Pt1000 temperature probe
automatic with Pt1000 or manual
200 values
4 ½ digit/character, 12 mm LCD display, shows
readings, °C & messages
6 tactile membrane keys
4...40°C
0...90 % (non-condensing!)
4 NiMH batteries, size AA
AC adapter for 230 V: model A4070
AC adapter for 115 V: model A4071
IP65 cabinet
195 x 101 x 44 mm
400 g
.
-1
Specifications
Keyboard
Error codes