Manganese HR with VARIO Powder Pack
0.1 – 18 mg/l Mn
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add the contents of one VARIO Manganese Citrate Buffer F10
Powder Pack straight from the foil into the water sample.
Close the vial tightly with the cap and swirl several times to mix the
contents.
Add the contents of one VARIO Sodium periodate F10 Powder
Pack straight from the foil into the same water sample.
Close the vial tightly with the cap and invert several times to mix the
contents.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 27)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Manganese.
Tolerance:
0.1 – 10 mg/l ± 0.2 mg/l
> 10 – 18 mg/l ± 0.5 mg/l