Factory setting

The ETS-D6 is supplied with the following factory settings:
Maximum temperature
Mode
pH-measurement
Controller
pH-Calibration
PT 1000-Calibration
Extension cable
Time Out Er 5

Language

Configuration

The unit is configured using the Setup menu:
Switch on the device
¢ LCD-Display: 1. Intro, see fig. 5
Press the SET button during the two second period.
Use the L and M buttons to navigate through the menu and sel-
ect the desired menu item.
The SET button is used to activate a selected menu item and con-
firm changes, see figs. 7, 8, 9.
The currently active settings are shown under the corresponding
menu item.
012007
450 °C
A
disabled
PID
OK
-
-
3 min
English
2. Information about the operating mode/
Maximum temperature and the request
"OK to enter Setup"
(Duration: 2 seconds), see fig. 6
Select the menu item "Exit" to leave the menu.
To accept and store changes select the "Yes" option when promp-
ted with "Confirm changes" on the next screen.
¢ LCD display see fig. 10
• Language
The following languages can be selected under the menu item
"Language":
- English
- German
- French
- Spanish
- Chinese
- Japanese
- Korean
¢ LCD display see fig. 11
Accept and store the selected language:
Press button SET
• Operating modes
The operating mode can be set under the menu item "Mode":
The ETS-D6 can be run in three different operating modes (A, B, C).
The mode and maximum temperature currently set will be
displayed for two seconds after power on, see fig. 6.
Mode A
The set-point temperature is always set to -10°C.
The maximum temperature can only be adjusted if Mode A is
active.
Use the L and M buttons to change the target temperature.
Mode B
The last target temperature selected will be used.
Use the L and M buttons to change the target temperature
19