Wiring Possibilities (Sensor Connection)

SensoTrans P 32200 P0/...
Sensor
Type
Strain
-7500 ... 7500mV/V
gauge
-7500 ... 7500mV/V
-7500 ... 7500mV/V

Connection for Strain Gauge Bridges (DMS)

13
V
Diff
Strain gauge
Terminal 1: Bridge supply voltage (+)
Terminal 4: Bridge supply voltage (-)
Terminal 2: Measured signal (+)
Terminal 3: Measured signal (-)
You can confi gure all functions of the transmitter using the Paraly SW 111 software.
Wirings with a shaded number can only be adjusted via IrDA.
Connection
Internal supply
External supply
Auto
1
2
3
4
14
=
Strain gauge
External supply (1 ... 3 V)
Terminal 1: Sense line (+)
Terminal 4: Sense line (-)
Terminal 2: Measured signal (+)
Terminal 3: Measured signal (-)
Adjustable via:
Figure
IrDA
13
x
14
x
13, 14
x
1
V
Diff
2
3
4
Switch
x
39