EMC data
Guidance and manufacturer´s declaration – electromagnetic emission - for all
EQUIPMENT AND SYSTEMS
The model BW-MTX1 is intended for use in the electromagnetic environments specified
below. the customer or the user of the model BW-MTX1 should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
ans voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
106
IEC 60601
test level
+/- 6 kV, +/- 8 kV contact
+/- 2 kV, +/- 4 kV, +/- 8 kV
+/-15 kV air
+/- 2 kV for power supply
lines
+/- 1 kV for Inpout/output
lines
+/- 1 kV line to line
+/- 2 kV line to earth
<5% U
(>95% dip in U
T
0.5 cycle
<5% U
(>95% dip in U
T
for 1 cycle
>70% U
(30% dip in U
T
25/30 cycles
<5% U
(>95% dip in U
T
for 5/6 sec
3 A/m,
30 A/m
BewellConnect - BW-MTX1 - User Manual - 2018-06_V1
Compliance level
+/- 6 kV, +/- 8 kV
contact +/- 2 kV,
+/- 4 kV, +/- 8 kV
+/-15 kV air
Not applicable
Not applicable
) for
T
)
T
) for
Not applicable
T
)
T
3 A/m,
30 A/m
Electromagnetic
environment -
guidance
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative humidity
should be at least
30%
Not applicable
Not applicable
Not applicable
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment