Installation; 4�1 Installation Location - Autosen AS001 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 Installation

Using process adapters the unit can be adapted to different process connections�
• Adapters have to be ordered separately as accessories�
4.1 Installation location
General
• The sensor tip is to be completely
surrounded by the medium�
• Immersion depth of the sensor: mini-
mum 12 mm�
Recommended
• For horizontal pipes: mounting from
the side�
• For vertical pipes: mounting in the rising
pipe�
Conditional
• Horizontal pipe / mounting from the
bottom: if the pipe is free from build-up�
• Horizontal pipe /mounting from the
top: if the pipe is completely filled with
medium�
To be avoided
• The sensor tip must not be in contact
with the pipe wall�
• Do not mount in downpipes that are
open at the bottom!