GENERAL INSTALLATION :

** Please follow the instructions below carefully to insure correct function of this device.
1. Insert the fibre optic probe into the handpiece and ensure that it is securely installed
at the fixed position (2mm of the fiber optic metal colour will remain visible).
2. Place the orange anti-glare cone onto the fibre optic probe and adjust the cone
position according to the dentist's preference.
3. Connect the DC adapter to the charger stand and plug the adapter into the mains.
** NOTE: Check to make sure the RED led light is lit and immediately switch to GREEN
led light at this point.
4. Place the handpiece in the charger stand to begin the initial charging.
(As shown in picture below)
** IMPORTANT :
> Please make sure the initial charge of the battery
is for at least 6 hours before the FIRST use.
This will insure the maximum length of the
battery life.
> When battery power decreases to the lowest
supply level, a continuous beeping sound will ring
for one minute to remind the user to recharge.
> Please do not charge the unit permanently.