Safety Precautions - Henry Schein HS-LED 1500 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
2

SAFETY PRECAUTIONS :

Please follow instruction steps in this manual for correct and safe operations of the
device. Pay special attention to the following warning signs beside the operation
descriptions where applicable.
a) DANGER
• This message appears where the operation may potentially cause severe
injury if instruction not followed correctly.
b) WARNING
• This message appears where the operation may cause serious damage or
defect to other objects.
c) CAUTION
• This message appears where the operation may cause slight damage to other
objects or slight bodily harm if operational instructions are not followed correctly.
d) IMPORTANT
• This message appears where the operation may involve risk of defect to the
device if not instructions are not followed correctly.
Sterilization of the Fibre-Optic Probe:
Before first use and after use on each patient, the fibre-optic probe should be disinfec-
ted or sterilized. It is recommended that the Probe be autoclaved at 134°C for 5 min-
utes. Suggestion is to use autoclave at 134°C for maximum results. The sterilization time
at 134°C should be 5 minutes.
Note: Please disconnect the Anti glare cone (orange) from the handpiece and
autoclave separately.
Product Cleaning:
If there is any dirt or stains on the surface of the control panel or surface of the hand-
piece, use a cloth to wipe clean. You can also use a neutral cleansing agent.
Note: Never use a solvent or volatile oil for cleansing as this can lead to fading
of the markings on the control panel.

Publicité

Table des Matières
loading

Table des Matières