Dc, Ac+Dc Current Measurement And E 4-20Ma% Reading - HT HT64 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

4.4.8. DC, AC+DC Current measurement and e 4-20mA% reading

Maximum input DC current is 10A (input 10A) or 600mA (input mAA). Do
not measure currents exceeding the limits given in this manual. Exceeding
voltage limits could result in electrical shocks to the user and damage to the
instrument.
Fig. 29: Use of the instrument for DC current measurement and 4-20mA% reading.
1. Cut off power supply from the circuit to be measured.
2. Select position A
20mA% for 4-20mA% reading.
3. Insert the red cable into input terminal 10A or into input terminal mAA and the black
cable into input terminal COM.
4. Connect the red lead and the black lead in series to the circuit whose current you want
to measure, respecting polarity and current direction (see Fig. 29).
5. Supply the circuit to be measured.
6. The value of DC current (see Fig. 29
7. The value of reading 4-20mA% (0mA = -25%, 4mA = 0%, 20mA = 100% and 24mA =
125%) (see Fig. 29
this function.
8. If the display shows the message
reached.
9. When symbol "-" appears on the instrument's display, it means that current has the
opposite direction with respect to the connection in Fig. 29.
10. To use the HOLD, RANGE and REL functions, see § 4.2.
11. For AC+DC measurement, see § 4.3.2 and to use the internal functions, see § 4.3
, mA
or 10A
left side) appears on the display.
right side) appears on the display. The bargraph is not active in
O.L
"
EN - 24
CAUTION
to measure DC current or the same position 4-
, the maximum measurable value has been
"
HT64

Publicité

Table des Matières
loading

Table des Matières