Calibration procedure (continued)
6.5

Calibration procedure notes

Prerequisites
The calibration device is in place or the programming
inputs are connected.
The transducer is connected to the system controller.
Voltage or current values from the transducer can be
read (using a multimeter or the system controller).
Values for null and end point
Any desired position of the magnet can be used as the
null or end point. However, the null and end points may
not be reversed.
The absolute null and end points must lie within the
minimum or maximum limits of what can be output
(see value table).
The distance between the null point and end point
must be at least 4 mm.
The last set values are always saved, regardless
of whether the setting was ended using the
buttons, the programming inputs or
automatically after 10 min have expired.
Value table for teach-in and adjustment
The following examples refer to transducers
with 0 to 10 V or 4 to 20 mA output.
For all other versions, use the values in the
value table below.
Output
Linear
gradient
transducer
Rising

BTL7-A...

BTL7-G...

BTL7-C...

BTL7-E...

Falling
BTL7-A...
BTL7-G...
BTL7-C...
BTL7-E...
Tab. 6-1:
Value table for teach-in and adjustment
www.balluff.com
Null point
Unit
Min.
Null
value
value
V
−0.5
0
V
−10.5
−10.0
mA
0
0
mA
3.6
4.0
V
+10.5
+10.0
V
+10.5
+10.0
mA
20.4
20.0
mA
20.4
20.0
Indication for
Indication
adjustment
for teach-in
2.0
4.0
2.0
4.0
6.0
12.0
6.0
12.0
8.0
6.0
−2.0
−4.0
14.0
8.0
14.0
8.0
Ignition protection
"nA" and "tb"
End point
End
Max.
Error
value
value
value
+10.0
+10.5
+10.5
+10.0
+10.5
+10.5
20.0
20.4
20.4
20.0
20.4
3.6
0
−0.5
−0.5
−10.0
−10.5
−10.5
0
0
20.4
4.0
3.6
3.6
english