BIOM® ready
Troubleshooting
Caution
Risk to the patient if the BIOM® ready is faulty.
If a fault occurs that you cannot rectify by following the
instructions below, dispose of the BIOM® ready in accordance
with the hospital's hygiene regulations and legal regulations
Fault
Possible Cause
Adapter or dovetail
Locking screw is too loose
mount is wobbling
Screws are loose
Cropped or decentered
The SDI®, other components, or the BIOM® ready adapter have
image
been mounted skewed
Dull image
Optical surfaces are dirty
Optical surfaces have been damaged by mechanical means
The front lens is touching the patient's eye
The patient's cornea is too dry
Blurred image
The BIOM® ready is maladjusted
The BIOM® ready model is not compatible with the microscope's
lens
Insufficient fundus ob-
The front lens is too far away from the patient's eye
servation
The microscope magnification has been set too high
Strong reflections on
The microscope illumination is switched on
the patient's eye or
lens
en
Instruction Manual
Remedy
Tighten the locking screw hand-tight
Tighten the screws with a suitable screwdriver
Re-mount correctly
Clean the optical surfaces
Select the correct work setting, clean the front lens
Wet the cornea regularly with a suitable solution
Focus the BIOM® ready in accordance with the instructions
Check the BIOM® ready model based on the reduction lens and if applicable, use the
other BIOM® ready model
Carefully reduce the distance with the microscope's focus
Reduce the microscope's magnification
Switch off the illumination; use intraocular illumination only
G/54413/0718/in