7. Grease the cleaned parts with lubricant that comes with the pipette.
8. Assemble the parts in reverse order.
9. Install the locking ring by pressing it with the screwdriver.
1-5 ml and 2-10 ml modules:
1. Detach the tip cone module by releasing the latch and removing the module.
2. Pull down the tip ejector as far as it goes. Detach the lower part of the tip ejector. (Snap fit). The upper
part can be slid upwards and removed.
3. Press the snap joints of the cylinder to release them. Pull out the cylinder.
4. Pull out the piston.
5. The piston assembly cannot be disassembled; only the o-ring can be replaced
6. Clean the cylinder, piston assembly and tip cone with a dry nap-free cloth.
7. Grease the cleaned parts with the lubricant that comes with the pipette.
8. Assemble the parts in reverse order.
Thermo Electron service department should handle all other service.
CAUTION!
If you send the pipette to us or to our local representative for service, please, ensure that the pipette has been
decontaminated before returning.
Also, please note that the postal authorities in your country may restrict the shipment of contaminated material by
mail.
TROUBLE SHOOTING
The troubleshooting guide below lists possible problems, causes and solutions.
Problem
Leakage
Inaccurate dispensing
Display is blank
CALIBRATE text on the
display and the pipette
won't operate
Pipette won't operate
Possible cause
Tip incorrectly attached
Foreign bodies between tip and
tip cone
Foreign bodies between the piston,
O-ring and tip cone
Insufficient grease on tip cone and
O-ring
O-ring damaged
Incorrect operation
Tips attached incorrectly
Calibration altered, possibly by
misuse
Inappropriate calibration.
High viscosity liquids may require
recalibration.
Battery is discharged
Power is OFF
Pipette is in reset mode
Tip cone module is
improperly attached
Solution
Attach tips firmly
Clean tip cone module; attach new tips
Clean and grease O-ring and tip cone.
Use grease
Grease accordingly
Change the O-ring
Follow instructions carefully
Attach tips firmly
Mount the pipette in the recharge stand,
and make sure that the light beside the
display is on
Move the pipette slightly
Lock the latch and select the module.
Press the dual stop trigger to the second
stop.
Release the latch, attach module firmly
to the pipette and lock the latch; press
the trigger to the second stop.
9