PerkinElmer OMNI Bead Ruptor 96 Manuel D'utilisation page 10

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

The following fuses are required for the Bead Ruptor
The following fuses are required for the Bead Ruptor 96:
• 2 x 2AT 250V (230V)
• 2 x 2AT 250V (230V)
• 2 x 5AT 250V (115V)
• 2 x 5AT 250V (115V)
• Unplug power cord from the power cord port (2).
• Unplug power cord from the power cord port (2).
• By pressing the locking device on the bottom side of the fuse compartment (1), fuse
• By pressing the locking device on the bottom side of the fuse compartment (1),
holder is released and you can pull it out.
holder is released and you can pull it out.
• Replace fuses.
• Replace fuses.
• Insert fuse holder and push it, until it locks.
• Insert fuse holder and push it, until it locks.
Power Switch is not illuminated when the Bead
RuptorTM 96 Homogenizer is on.
Display screen is not working
Power Switch is not illuminated
when the Bead Ruptor 96 is on.
"LID" appears on display
Display screen is not working
Message "Er1" appears on display screen
"LID" appears on display
Message "Er2" appears on display screen
Message "Er1" appears on
display screen
Message "Er2" appears on
Message "Er3" appears on display screen
display screen
Message "Er3" appears on
DO NOT: attempt to service the Bead Ruptor
display screen
than those discussed in this manual. For any issue that is unsuccessfully corrected
using this guide, please contact your authorized dealer or call
DO NOT: attempt to service the Bead Ruptor 96 in a manner other than
OMNI International, a PerkinElmer company at 1-800-776-4431.
those discussed in this manual. For any issue that is unsuccessfully correct-
Replacing the Fuse
Replacing the Fuse
Troubleshooting
Troubleshooting
Error
Error
96 Homogenizer:
TM
1. Check the power supply
2. Check fuses and replace if necessary
3. Call for service
Call for service
1. Check the power supply
Indicates that the lid is open when the start
2. Check fuses and replace if necessary
button was pressed. Close lid and clear mes-
3. Call for service
sage by pressing the Start/Stop button. Press
Start/Stop to begin cycle.
Call for service
Motor has over heated. Turn power off and
Indicates that the hood is open when the start button
allow motor to cool down.
was pressed. Close hood and clear message by pressing
The Bead Ruptor
the Start/Stop button. Press Start/Stop to begin cycle.
to reach the set vibrational frequency. Check
Motor has over heated. Turn power off and allow motor
Milling Jars for obstructions. Switch Power
to cool down.
OFF and ON to clear error message.
The Bead Ruptor 96 is unable to reach the set vibrational
Lid was opened during operation. The power
frequency. Check Milling Jars for obstructions. Switch
must be turned OFF and then ON to clear
error message before operation can continue.
Power OFF and ON to clear error message.
Hood was opened during operation. The power must be
96 Homogenizer in a manner other
TM
turned OFF and then ON to clear error message before
operation can continue.
1. Fuse Compartment
2. Power Cord Port
Possible Solution
Possible Solution
96 Homogenizer is unable
TM

Publicité