Space Requirement; Connection To The Mains; Commissioning; Opening And Closing The Lid - Hettich HAEMATOKRIT 210 Mode D'emploi

Centrifugeuse de laboratoire
Masquer les pouces Voir aussi pour HAEMATOKRIT 210:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
6

Space requirement

According to the laboratory instrument standards EN 61010-2-20 an emergency switch to separate power supply
in the event of a failure must be installed in the building electrical system.
This switch has to be placed remote from the centrifuge, preferred outside of the room in which the centrifuge is
installed or near by the exit of this room.
The necessary space requirement can be found under Dimensions in the Technical data chapter.
The centrifuge must be set up in a suitable place, so that it is stable. During set up the required safety area of
300 mm around the centrifuge, in accordance with IEC 1010-2-2, must be observed.
Persons and hazardous materials must not be located in the safety area whilst the centrifuge is in
operation.
Substances can be discharged through the ventilation opening on the rear side of the centrifuge. The device is to
be placed in such a way that the air stream is not directed towards people.
Do not place any object in front of the ventiduct.
− Keep a ventilation area of 300 mm around the ventiduct.
7

Connection to the mains

Check whether the supply voltage, supply frequency and on-site mains fuse agree with the specification on the
nameplate. The nameplate is located on the back of the centrifuge.
Make sure that the mains switch is in the "0" position.
The centrifuge must be connected to a standard mains socket using the power supply cable provided.
8

Commissioning

Check that the centrifuge has been properly set up and that the electrical connections are correct (see
Connection to the Mains and Space Requirement).
Switch the mains switch "ON", switch position "I".
After a short time the control panel will switch on and the last set parameters will appear in the displays.
When the
symbol has lit up, the lid can be opened.
Open the lid and remove the transport protection (see Transport Protection information sheet).
9

Opening and closing the lid

9.1

Opening the lid

When the
symbol has lit up, the lid can be unlocked and opened.
Unlock the lid by pressing the
The
symbol appears.
The lid can only be opened if the centrifuge is switched on and the rotor is stationary.
If this is not possible, see chapter " Emergency release ".
9.2

Closing the lid

Close the lid by swinging the lid down and pressing lightly. If the symbol
has been correctly locked.
Do not bang the lid shut.
22/58
key and swivel the lid upwards.
lights up in the rotation display the lid