Rs 232 Communication Port - Varian PTB Turbo V 700 HT 969 9445 Notice D'emploi

Masquer les pouces Voir aussi pour PTB Turbo V 700 HT 969 9445:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
Figure 16

RS 232 Communication Port

Communication serial port connections and mini- mum connection
configuration are shown in the following figures. The communication
port mating connector is supplied with the RS 232 PCB
(AMP/Cannon or equivalent 9-pin "D" type male connector). The
external cable (not supplied) between the host computer and the
controller requires some crossed wires so that signals are connected
correctly.
For example, the Transmit data signal from controller (pin 2) must be
connected to the host computer's Receive data line (pin 3) and vice
versa. Consult the host computer's instruction manual for its serial
port connections.
Agilent cannot guarantee compliance with FCC regulations for radiated
NOTE
emissions unless all external wiring is shielded, with the shield being
terminated to the metal shroud on the O-subconnector. The cable should be
secured to the connector with screws.
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
Serial port connector installation