Filter
Applies a data acquisition filter (Main filter or Pre-filter). The user can select the filter
parameters
parameters.
Main filter
• Fixed Period Averaging (FPA)—Fixed Period Averaging averages data over a user
• RCA time constant—The RC filter helps smooth out turbulence through the use of a
Pre-filter
• Median filter—The filtering process is done in the sensor. The feature can be disabled.
Wet/dry
Sets the sensor submersion threshold for wet and dry conditions. The default value is 20%.
threshold
Refer to the appendix in the expanded version of the manual for more information.
Auto zero
Sets the Auto Zero feature to On or Off.
depth
If set to On, the instrument does an air calibration when the sensor is removed from the
water and is in the air. To do the air calibration, the instrument automatically zeroes the
sensor.
If set to Off, the user can manually zero the sensor. To do this, remove the sensor from the
flow, then push and hold the Quick Jump key for five seconds.
When the sensor has been in the flow for 30 minutes, the green circle in the upper right
corner goes from green to red. This is a prompt to the user to remove and zero the sensor
again.
EMI
Sets the local line frequency for ambient noise rejection to 50 Hz (default) or 60 Hz .
Clock
Sets the date and time of the portable meter in 24-hour format. Daylight savings time is not
supported.
USB
Sets the USB mode.
• Mass Storage (default)—This mode operates like a memory stick or hard drive. Files are
• CDC—This mode is used to update firmware.
Language
Selects the language used in the menus.
Units
Sets the units for velocity, flow and depth measurements. Options are Metric (default) or
English .
Beeper
On (default) or Off. If set to On, the meter makes an audible tone when:
• The sensor is at the correct depth for applicable profile methods.
• 95% of the measurement time is complete (profiler mode).
• An inactive button is pushed in any menu.
This feature is available only with the optional depth sensor.
Flow
Selects the method of flow calculation for open water segment (stream profiles only).
calculation
Options are Mean-section or Mid-section. Refer to the appendix in the expanded version of
the manual for more information.
18 English
selectable fixed period of time (1 to 480 seconds). The default is 30 seconds. If the FPA
value is 5, the velocity value shown in the display is updated once every 5 seconds.
selected time constant in the filter algorithm. This mode is useful when searching for a
maximum velocity, for example in the common 0.9 x Vmax profile method. High RC filter
time constants give higher degrees of smoothing. The time constant can be set from 2 to
20 seconds, with a default value of 6. At 1 time constant, the filter settles to
approximately 60% of the final value. At 5 time constants, the filter settles to 99.9% of
the final value. Thus, if the RC value is set to 2, the final value shows after 10 seconds.
However, the recommended (default) value is 5. Enable the feature to enter or change
this value.
read-only.