Disassembly
The disassembly of the set must be performed also without additional tools and occurs in the reverse order of
the assembly:
Hold the Arm support with one hand and loosen with the other hand the fastening screw of the rail clamping
base. Remove the base from the rail of the surgical table.
Handling of the articulated arm
A
EN
1
Hold the articulated arm with one hand in the anterior segment (A) and
manipulate the central clamping handle (Z) with the other hand
2
To release turn the central handle (Z) counter clockwise (B) as far as
necessary
Move the articulated arm into the desired position
3
4
To secure, turn the central handle (Z) clockwise (C)
5
Check that the articulated arm is fixed firmly and is functioning properly
DANGER!
The articulated arm can move unintentionally if the central clamping is released.
Always hold the arm board with one hand and manipulate the central clamping
device (Z) with the other hand.
DANGER!
If the articulated arm is not tightened properly it can become loose and move,
potentially causing injuries.
The articulated arm must be tightened properly.
DANGER!
A damaged product can cause injuries.
Use only products in proper condition and check functionality.
DANGER!
Threat through inappropriate handling
Follow instructions for use of arm support and surgical table
WARNING!
Please check the risk of collision during patient transport with exposed arm
support.
If possible position the arm support parallel to the surgical table.
WARNING!
The arm support lateral 4646.XX has a maximal carrying capacity of 6 kg
during intended use, the arm support lateral 6646.XX one of 10 kg, the arm
support vertical 6656.XX has a maximal carrying capacity of 6 kg and the arm
support vertical 4670.XX has a maximal carrying capacity of 4 kg.
The maximal carrying capacity must not be exceeded.
20