Guidance and manufacturer's declaration – electromagnetic emission
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601-1-2
Immunity test
test level
Electrostatic
±2kV, ±4kV,
discharge (ESD)
±8kV, ±15kV air
EN 61000-4-2
±8kV contact
Compliance
Guidance
Group 1
This equipment uses RF energy only for its internal
functions. Therefore its RF emissions are very low
Class A
and are not likely to cause any interference in nearby
electronic equipment.
This equipment is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply
Class A
network that supplies buildings used for domestic
purposes.
Complies
Compliance level
±2kV, ±4kV, ±8kV, ±15kV air
±8kV contact
Electromagnetic
environment-guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material the relative
humidity level should be at
least 30%.