G
Malfunction
Display Er- 300
Display Err -1
Display Err -2
Display 0
0
Measurement interrupted and
cuff deflated and re-inflated
GB-50
Possible reasons
Excessive cuff pressure
The cuff inflated to the maximum by
movement of the arm or body
Measurement error
Measurement could not be carried out
due to movement of arm or body
Inflation error
Pressure built up too quickly or too
slowly
Error was recognized
Defective measurement
Start/stop button accidentally pressed
while changing battery
Device detects defect or diastolic value
too low
Movement during measurement
Appendix
Corrective action
Repeat measurement, no moving the
arm or speaking
Repeat measurement, no moving the
arm or speaking
Check position of air plug, repeat
measurement
Repeat measurement, no moving the
arm or speaking
Reboot device with Start/stop button
Device shows no defect
Repeat measurement
Interrupt measurement if necessary
New measurement after 5 minutes
break