Technical Specification - Arjohuntleigh Flowtron Excel Notice D'utilisation

Masquer les pouces Voir aussi pour Flowtron Excel:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 78

11. Technical Specification

Pump
Model:
Part Numbers:
Supply Voltage:
Supply Frequency:
Power Input:
Size:
Weight:
Case Material:
Plug Fuse Rating:
Pump Fuse Rating:
Degree of protection
against electric shock:
Degree of protection
against liquid ingress:
Mode of operation:
Cycle or Therapy Modes:
Pressure Range:
ENVIRONMENTAL INFORMATION
Condition
Operating
Storage and Transport
(Long Term)
Storage and Transport
(Short Term)
Products should be well packaged and stored in a well-ventilated, dry and non-corrosive environment.
Flowtron Excel
UK/AUS
247003 - UK
247003AU - AUS
247004 (247EUR) - EU
230 V
50 Hz
35 VA MAX
133 x 152 x 275 mm
2.7 kg
Fire Retardant ABS Plastic
5A BS1362 (UK ONLY)
T1AL 250 V
Class II, Double Insulated with Functional Earth
Type BF
IPX0 - No protection
Continuous
60 seconds total
12 seconds inflation
48 seconds deflation
30 - 60 mmHg (±4 mmHg)
Temperature Range
+10 °C to +40 °C
(+50 °F to +104 °F)
+10 °C to +40 °C
(+50 °F to +104 °F)
-20 °C to +50 °C
(-4 °F to +122 °F)
If the pump is stored in conditions outside of the
"Operating" ranges, it should be allowed time to
stabilise at normal operating conditions before use.
14
KSA
247009KSA
230 V
60 Hz
35 VA MAX
Relative Humidity
30% to 75%
(non-condensing)
20% to 95%
(non-condensing)
20% to 95%
(non-condensing)
USA
247001
120 V
60 Hz
35 VA MAX
Atmospheric Pressure
700 hPa to 1060 hPa
700 hPa to 1060 hPa
500 hPa to 1060 hPa

Publicité

Table des Matières
loading

Table des Matières