KERN and SOHN MRS-A01N Mode D'emploi page 8

Etalonnage de la température
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PRESS START KEY
TEMP 25°C
20.00MIN
TEMP 100°C
20.00MIN
MRS-A01N-BA-def-1112
To start temperature calibration press
«START STOP».
Temperature calibration has been
activated. The dryer is heated to the first
temperature value. The temperature and
the remaining temperature calibration
time are displayed in the information strip.
Heating to the first temperature value
takes approx. 20 minutes.
After the calibration temperature is
reached, the indication blinks. In order to
adjust, read the current temperature using
a thermometer and enter it. Select the
digit using the
increase/decrease the value using the
buttons. Afterwards press
«START STOP».
The device will then be heated to the
second temperature value.
After the calibration temperature is
reached, the indicator blinks again. Read
the current temperature using a
thermometer, enter it and confirm it using
the «START STOP» button.
If an optional printer is connected, a
protocol will be printed after the
calibration is complete .
button and
8