Calibrating/adjusting the temperature
We recommend periodic checks of device temperature using the optional
temperature calibrating set MRS-A01N. Temperature is measured in two points
(default setting 100°C and 160°C) – correction is possible in both of them.
Earlier, after the heating phase is complete, the device should be left too cool to
surrounding temperature.
1. Calibrating temperature in Configuration menu
In order to enable temperature calibration, the option "Temp. CAL." Must be
set to "ON" (see chapter. 4.4.6 „Temperature calibration"):
Turn on the humidity meter pressing «ON/OFF».
When turning the device on hold «MENU» (for about 10 seconds), until
information "SET DATA PRINT" is displayed erscheint.
Use the «
"OFF" setting will be displayed.
MRS-A01N-BA-def-1112
KERN MRS A01N
Version 1.0 11/2011
Operating Instructions
Temperature calibration set
» button to select the "Temp. CAL" option. Current "ON" or
6