KERN and SOHN MRS-A01N Mode D'emploi page 7

Etalonnage de la température
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TEMP. CAL.
TEMP. CAL.
In order to change the setting from "OFF" to "ON", use the following settings:
TEMP. CAL.
TEMP. CAL.
TEMP. CAL.
1. Temperature calibration/adjustment
TEMP. CALIBRATION
T100 T 160
START
7
ON/OFF
Temperature calibration activation
OFF
If the setting is "OFF"", press the «
button, the "OFF" indication will blink.
ON
If the setting is "ON"", press the « » button,
button, the "ON" indication will blink.
ON
Press the «
Press «esc» to leave the menu
» button to confirm
Make sure that drying is not performer –
that is that the device is In weighing
mode.
Insert the temperature sensor bard to
the sample chamber. Close the sample
chamber.
Press «TARE» until
„TEMP. CALIBRATION" setting is shown.
Calibrate according to the default values
to temperatures of 100°C and 160°C. To
set temperature calibration value to
individual demands, use the
(100°C) button or the
button. Use
requested temperature. Use the
button to pass to the next number and
enter the selected value. Then confirm the
entered values using the «START STOP»
button.
»
(160°C)
and
to set the
MRS-A01N-BA-def-1112