The air leaving the collection container must pass through a microbiologi-
cal filter before entering the suction device.
The associated equipment or medical devices connected to the suction
ejector must be fully compatible with the characteristics and perfor-
mance data of the MediEject II suction ejector.
* In some markets, it is recommended to use a safety container installed
between the vacuum source and the collection container before the ejec-
tor.

6.2. BEFORE USE

Visual Inspection before fitting
Check if there is visible external damage on the suction ejector and its
associated equipment (including product labels and marking). If it shows
sign of external damage, remove the products from service and identify
its status.
Visually check if the ejector and its associated equipment are contami-
nated; if needed, apply the cleaning procedure according to cleaning
procedure here under.
Check if the service or the disposal time has not been exceeded, using
GCE or owner's date coding system. If service or disposal time has been
exceeded, remove the product from service & suitably identify its status.
Fitting to medical gas pipeline system - terminal units/mobile gas source
Refer to appendix Nr. 2 to get information about connecting / disconnect-
ing procedure.
For ejector fitted with a rail clamp and a flexible hose, clip the suction eje
tor on the rail prior to connect the quick coupler.
Functional test before use and Leakage test
After connecting the medical pipe line system terminal unit/portable
gas source and associated equipment, turn the vacuum control knob to
maximum position (counter clockwise).
Set the suction ejector ON by pulling up the integrated handwheel.
Check that there is a significant vacuum level at patient side inlet by
locking the patient side inlet with your finger.
Set the suction ejector OFF by pushing the integrated handwheel and
check if vacuum level is without change at locked patient side inlet by
your finger.
Set vacuum level to minimum by turning the integrated handwheel it
clockwise.
Before connecting any associated equipment to the product, make sure
that the push-pull on/off valve is in the closed position and that the pa-
tient is not connected.
If any leaks are found, check the connection and repeat the leak test.