Complete Charging - Axonics SNM System 1401 Manuel D'utilisation

Masquer les pouces Voir aussi pour SNM System 1401:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 86
You will hear a set of rising tones repeat 3 times when charging is
1
complete. The Charger indicator light will turn off. This will take
1 to 4 hours for a low battery, depending on the implant depth
of your Stimulator and how well the Charger is aligned over the
Stimulator.
Note: If you hear 3 beeps and feel the Charger vibrate, charging was
interrupted before completion. See Troubleshooting for potential causes and
resolutions.
Note: It is safe to interrupt charging if you need. Partial charging is
acceptable if you do not have time to fully charge the Stimulator.
Remove the Charger from your skin by peeling off the Carrier or
2
unstrapping the Belt. Take the Charger off the Carrier or Belt.
Dispose of the Carrier or store the Belt. Place the Charger on the
3
Dock.
Note: The Charger should provide 5 or more years of service. With repeated
charging, the Charger battery may lose capacity or charge the Stimulator
more slowly. Notify your doctor if you experience a significant change in
the performance of your Charger. For example, notify your doctor if your
Charger takes over 6 hours to fully charge on the Dock, or if a fully charged
Charger cannot completely charge your Stimulator battery.