Routine Operation; Configuration; Interlocks; Battery Charge Level - ACteon Mini Led III SuperCharged Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
MLED III Supercharged U7
successive one second flashes with a rest of period of
250 ms between each flash) - (show 05);
- 10 shots of 1 second (full power with emission of 10
successive one second flashes with a rest of period of
250 ms between each flash) - (show 10) ;
- Step-Cure mode:
- 6 seconds progressively and 3 seconds at full power
(show 09) ;
- 10 seconds progressively and 5 seconds at full power
(show 15).
V – ROUTINE OPERATION
The Mini LED SuperCharged is normally placed on its
holder. The first time you plug it into the battery it will
activate in the Menu Fast-Cure and the exposure time of
3 seconds.
Your Mini LED SuperCharged is ready to operate once the
menu and the exposure time has been selected.
Place the light guide as close as possible to the surface
of the material to be polymerized, without touching it
as this may adversely affect the quality of the
polymerization. The use of a transparent strip is
recommended.
The polymerization starts by pushing on the ON/OFF
button. The setting will be confirmed by an audible
signal (beep).
A countdown on the LCD screen shows you how much
time remains.
When the polymerization cycle is completed, the time
that was applied is displayed.
You can press the ON/OFF button at any time to
interrupt the current polymerization cycle.
After 5 minutes without use, the unit goes into standby
mode (low consumption). The green indicator light and
the LCD screen backlighting switch off.
The unit can be taken out of standby mode by simply
pressing one of the three buttons (which will not
activate that function when coming out of standby
mode).
V6:Mise en page 1
05/09/2011
VI – CONFIGURATION
6.1 STARTUP
On startup, the unit performs an automatic test
sequence (auto-check).
The device automatically recalls the settings used for
the previous dental treatment.
6.2 MODES
The Fast, Pulse and Stepped modes are selected using
the left button (fig.2-6) on the handpiece.
6.3 EXPOSURE TIMES
The exposure times are selected by pressing the right
button (fig.2-7).
VII – INTERLOCKS
The Mini LED SuperCharged is equipped with a system
for the detection of possible unit operation anomalies.
7.1 EXCESSIVE TEMPERATURE
During intensive use, an excessive temperature may be
detected and the screen then displays the letters "OH"
(Over Heat), the audible warning (beep) sounds 4 times
and the indicator light (two-color LED) becomes red.
It is then recommended that the user leaves the
handpiece to cool for a few minutes, until the indicator
light becomes green again and the screen once again
displays the selected time cycle.

7.2 BATTERY CHARGE LEVEL

When the LCD screen displays the last remaining round
battery charge indicator, the audible alarm (beep)
sounds twice.
You then have about 500 seconds remaining before the
unit becomes completely discharged. It is then
recommended, if possible, to charge the battery for
further use.
When the battery becomes completely discharged, the
letters "Lb" (Low Battery) are displayed on the screen,
the audible alarm (beep) sounds 4 times and the
indicator light (two-color LED) becomes red.
11
09:53
Page 11

Publicité

Table des Matières
loading

Table des Matières