Inlet Screen Installation
The inlet screen mod. 9699138 prevents the blades of the pump from
being damaged by debris greater than 5.2 mm diameter.
The inlet screen, however, does reduce the pumping speed by about
10%.
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
It is welded to the center ring of the ISO 250 F viton gasket.