Télécharger Imprimer la page

Diagnostic Buffer - Agilent Technologies Turbo-V 2K-G Notice De Mode D'emploi

Publicité

4
5
6
F
The ProfiBus external diagnosis services are used to report following
unexpected situations:

Diagnostic Buffer

Offset
0
1
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Controller response was "Over range"
Controller response was "Bad operation"
Controller response was an unknown response
Parameter not supported by the controller
Controller fail: if the controller goes in fail mode, an external
diagnostic service is re-quired and one byte in the diagnostic
buffer signals the type of failure.
User parameter mismatch: if the user sets one or more
parameters out of their allowed range, a ProfiBus external
diagnostic service is required and some bits in the diagnosis
frame signals which parameter is wrong (i.e. over range or under
range).
Meaning
B
station non exist (set by master)
0
B
station not ready (slave not ready for data exchange)
1
B
cfg fault (configuration data doesn't match)
2
B
ext diag (slave has external diagnostic data)
3
B
not supported (slave doesn't support requested function)
4
B
invalid slave response ( slave sets permanent '0')
5
B
prm fault (wrong parameter assignment)
6
B
master_lock (slave is parameterised by another master)
7
B
prm req (slave has to be re-parameterised)
0
B
stat diag (static diagnosis)
1
B
fixed '1'
2
B
wd_on (response monitoring active)
3
B
Freeze mode (received freeze command)
4
B
Synch mode (received synch command)
5
B
reserved
6
B
diag deactivated (slave is parameterised by another master)
7
Technical Information
Profibus Message Mapping
(H)
17
343/360

Publicité

loading

Ce manuel est également adapté pour:

96988719698873