Télécharger Imprimer la page

J2 - Serial - Agilent Technologies Turbo-V 2K-G Notice De Mode D'emploi

Publicité

1=Remote-I/O
Pin-10 Low -> Purge Close
Pin-10 High -> Purge Open
Pin-12 Low ->Vent Close
Pin-12 High -> Vent Open
win.122=0 Vent Close (def)
win.122=1 Vent Open
win.145=0 Purge Close (def)
win.145=1 Purge Open
Fail -> Vent and Purge =
Closed
Power Fail -> Purge and
Vent = Closed
J2 – Serial
This is a 9 pin D-type serial input/output connector to control via an
RS 232 or RS 485 connection the Turbo-V 2K-G.
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Turbo-V 2K-G - Vent and Purge Operation
1=OnCommand
Win
146
0=Serial (def)
Purge
O
C
Vent
O
C
Stop
Technical Information
0=Auto (def)
Win
125
Win.126 set the "Vent Valve Opening Delay", from 0 to
65535, 1bit=0.2s, def=10s=50bit
Win.147 set the "Vent Valve Open Time", from 0 to
65535, 1bit=0.2s, def=0=infinite
Stop Command
Win.
126
Ramp
Normal AutoTun
Stop
Breaking
Start Command
(H)
17
Interconnections
Win.
147
If win.147=0
317/360