One-point calibration of temperature and relative humidity
CAL1: CAL1 (one-point calibration) is used to set an offset for sensor 1
(temperature). The offset value appears at the bottom of the display. The limit
values are +/- 10°C or +/- 10°F.
The selection is made using
and
; and entered with
.
To obtain the factory calibration, set the offset to 0.0.