• Make sure that the unit does not come in contact with water or other
liquids as long as the plug is in the power socket, especially if you
use it in the laboratory near water or liquids.
• Using a two-pin AC adapter / connecting cable is NOT recommended.
• Make sure that during operation the power cable (20) never gets wet
or damp. Locate the cable so that it is not trapped, jammed or
otherwise damaged. If the power cable (20) or the plug (19) is
damaged, the cable must first be replaced before using this device.
• Do not cover the ventilation grilles (10) and cooling fans (11) on the
back of the unit.
• Never insert plastic inoculating loops or objects with a temperature-
resistance of less than 1200ºC into the annealing tube (9, 9a).
• Never hold fingers or other parts of the body on or in the sterilization
!
orifice (9a). Danger of burns!
Surface
may be hot
• If the IR hologen lamp is illuminated, do not look directly into the
sterilization orifice (9a). Distance eye > 200 mm and distance skin
> 30 mm. Never operate the device without the annealing tube in
place (9)!
• Only use original quartz glass annealing tube.
• Only remove the annealing tube (9) after it has cooled down. Even
some time after use, the sterilization orifice (9a) and the annealing
tube (9) are still hot. Danger of burns! Allow the unit to cool (apt. 10 min.)
and shut it down before cleaning, disinfection, maintenance or transport.
• After cleaning, let the annealing tube (9) dry first before reinstalling.
• Do not handle the annealing tube with bare hands. Grease or other dirt
can be removed with a soft cloth with alcohol.
• After using the device or if it is unused for longer periods, turn off the
power at the power switch (17).
• The upper part of the glass front can become hot if the device is in
continuous use.
• In the event of faults or before opening the device, remove the mains
plug (19) from the power socket.
• Keep all equipment and items that can be damaged by a magnetic
field (such as credit & debit cards, data disks, mechanical watches,
etc.) well away from the magnetic surfaces of the glass front (M).
For example the minimum distance to the glass front (M):
credit & debit cards 32 mm, heart pacemaker 40 mm.