Insulation Resistance (Riso); No Trip And Hi Current Modes Rcd/Fi - Beha-Amprobe Telaris Série Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 68
Range R
LO
7.5 Ω
35 Ω
240 Ω
2000 Ω
Test Probe Zeroing
Live Circuit Detection
Insulation Resistance Measurement (R
Test Voltages
Accuracy of Test Voltage
(at rated test current)
Test
Insulation
Voltage
Resistance Range
100 V
100 kΩ to 20 MΩ
20 MΩ to 100 MΩ
250 V
10 kΩ to 20 MΩ
20 MΩ to 200 MΩ
500 V
10 kΩ to 20 MΩ
20 MΩ to 200 MΩ
200 MΩ to 500 MΩ
1000 V
100 kΩ to 200 MΩ
200 MΩ to 1000 MΩ
Note: The number of possible insulation tests with a fresh set of batteries is 1750.
Auto Discharge
Live Circuit Detection
Maximum Capacitive Load

No Trip and Hi Current Modes RCD/FI

Mains Input Voltage Range
Input Connection
(soft key selection)
Limit on Consecutive Tests
Maximum Test Current @ 400 V 12 A sinusoidal for 10 ms
Maximum Test Current @ 230 V 7 A sinusoidal for 10 ms
Test Current
210 mA
100 mA
20 mA
2 mA
Press the F3 to compensate the test probe.
Can subtract up to 2 Ω of lead resistance.
Error message for >2 Ω.
Inhibits test if terminal voltage >10 V ac detected prior to
initiation of test.
)
ISO
100-250-500-1000 V
+10 %, -0 %
Resolution
0.01 MΩ
0.1 MΩ
0.01 MΩ
0.1 MΩ
0.01 MΩ
0.1 MΩ
1 MΩ
0.1 MΩ
1 MΩ
Discharge time constant <0.5 second for C = 1 μF
or less.
Inhibits test if terminal voltage >30 V prior to
Operable with up the 5 μF load.
100 - 500 V ac (50/60 Hz)
Loop Impedance: phase to earth
Line impedance: phase to neutral
Automatic shutdown when internal components are too
hot. There is also a thermal shutdown for RCD tests.
26
Test Current
Accuracy
1 mA @ 100 kΩ
±(5 % + 5 digits)
±(5 % + 5 digits)
1 mA @ 250 kΩ
±(5 % + 5 digits)
±(5 % + 5 digits)
1 mA @ 500 kΩ
±(5 % + 5 digits)
±(5 % + 5 digits)
±10 %
1 mA @ 1 MΩ
±(5 % + 5 digits)
±10 %