Technical Information
Interconnections
FRG-700 measure from 5x10
atmosphere) and its temperature range goes from 5 °C to 55 °C
[41 °-131 °F].
Pressure data is available on three independent reading channels:
1.
2.
3.
Pressure gauge can be connected/disconnected from the TPS-
compact during normal operation.
Figure 21
The Mini-IMG gauge is intended for high vacuum use, so with pressure above
NOTE
50 mTorr the signal is misinterpreted and a "false" pressure is displayed. Use a
transition gauge to understand if the reading is really under 50 m Torr or not.
224/262
On-board display (data expressed in mbar, Torr or Pa) – (see
Pressure Display Module picture)
Serial communication line (Win 224)
Analogical output.
Gauge Connector
-9
mbar to atmosphere (3.8x10
TPS-compact User Manual / 87-901-010-01
-9
Torr to