Hach SC4500 Manuel D'utilisation De Base page 17

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 77
The power cable plug is used to connect and disconnect power to the controller. For installation in
conduit, the installed local disconnect is used to connect and disconnect power to the controller.
Refer to
Figure 9
and
appropriate terminal until the insulation is seated against the connector with no bare wire exposed.
Tug gently after insertion to make sure that there is a secure connection. If necessary, remove the
connector from the PCBA for easier wiring of the terminals.
Note: Make sure that all of the cables stay below the cable limit line printed on the PCBA to prevent interferences
with the high-voltage barrier. Refer to
Figure 9 Connect conduit or a power cord
1 AC and DC power terminal
2 Cables limit: do not put cables above the
line.
Terminal
Description
L
N
Protective earth ground
Terminal
Description
L
N
24 VDC return
Protective earth ground
Table 2
or
Table 3
to connect conduit or a power cord. Insert each wire into the
Figure
9.
Table 2 Wiring information—AC power
Hot (Line 1)
Neutral (N)
Table 3 Wiring information—DC power
+24 VDC
3 Protective earth ground
4 Conduit hub (or strain relief fitting for power
cord)
Color—North America
Black
White
Green
Color—North America
Red
Black
Green
Color—EU
Brown
Blue
Green with yellow stripe
Color—EU
Red
Black
Green with yellow stripe
English 17

Publicité

Table des Matières
loading

Table des Matières