1. INTENDED USE

Vibracap R&S is a universal mixer for amalgam, glass ionomer cements and other pre-dosed dental materials in capsules and
syringes.
The apparatus has been developed solely for use in dentistry. Liability cannot be accepted for damage resulting from misuse
or failure to observe the Operating Instructions.
The user is responsible for testing the Vibracap R&S for its use and suitability for the intended purposes. The intended use
also includes the observation of the notes and regulations in these Operating Instructions. The Vibracap R&S has been
shipped from the manufacturer in a safe and technically sound condition.
The apparatus is built according to: Directive 93/42/EEC and subsequent amendments contained in according with the
Directive Rule 89/336/CEE, 92/31/CEE and 98/68/CEE and the following harmonized standards have been applied: EN 61010-
1 - EN61010-2-81-EN61326-1 - EN61000-3-2 – EN6100-3-3 - ISO 7488.

2. SAFETY

WARNING
CAUTION!
Please read these instructions carefully before you set up and start the device. As with all technical equipment, this unit will
function properly and be safe to operate only if both the general safety regulations and the specific safety instructions laid
out in these Operating Instructions are followed.
1. The unit must be operated accordance with the following instructions. We accept no liability for any damage arising from
use of the unit for any other purpose.
2. Prior to setting-up the device, ensure that the voltage stated on the rating plate is identical to the available main voltage.
Use of the device at different voltages may lead to the destruction of the unit.
3. The unit must be operated with the cover intact and closed, otherwise, there is a risk of injury.
4. The cover must not be opened until the capsule holder has come to a complete stop.
5. Only the capsule to be mixed but no other objects may be present in the space below the cover prior to initiating the
mixing cycle.
6. Placement or operation of the unit on non-level surfaces is not permissible.
7. For operation of the unit, connect the main plug to an adequate electrical socket with grounded lead. If an extension cable
is used, ensure that the grounded lead is not interrupted.
8. In order to prevent electrical shock hazards do not insert any objects into the unit with the exception of replacement parts
properly installed in accordance with the present instructions.
9. The device should be opened, and repairs be carried out only by qualified personnel.
10. Only adequately labelled original Vibracap R&S replacement parts may be used in the repair and maintenance of this unit.
Any damage resulting from the use of third-party replacement parts voids the liability of the manufacturer.
11. Do not expose the device to solvents, flammable liquids, and strong heat sources as these may damage the plastic
housing.

3. DENTAL MATERIALS USED

The corresponding instructions for use of the material suppliers have to be observed.

4. ASSUMPTION OF IMPAIRED SAFETY

If it has to be assumed that safe operation is no longer possible, the power must be disconnected to avoid accidental
operation. This may be the case, for example, if the apparatus is visibly damaged or no longer works correctly. A complete
disconnection from the power supply is only ensured when the power cord is disconnected from the power source.
10