4: Installation AXONIC with moving flanges
o
4 - ELECTRICAL CONNECTION AND POWER SUPPLY
AXONIC must be power-supplied by the integrator (CF 51, CF 55 or CF 800) or an exterior power supply. (eg: Itron Pulse-
Box or other integrator with galvanic insulation).
4.1 - Connecting the flow meter to integrator unit
Beside the main pulse output A the product could offer (as option) a 2nd pulse output B providing pulses and backflow to
another independent device (e.g regulator). This option would be configured at factory.
This pulse output B is following the same hardware specification as output A. Pulse-weight and pulse length of output A and
B are independent.
Before connecting, make sure the pulse weights of flow meter and integrator are equal (typeplate of both devices)!
Connect to integrator according to the following connecting diagram (figures show connecting diagram for CF 51, CF 55,
Pulse-Box and CF 800).
Before connecting an Axonic to other Systems to other systems make sure that a galvanic insulation is guaranteed.
(e.g. by using the Itron Pulse-Box)
 
 
Fig N
5: Mounting of an Itron integrator CF51/CF55 onto AXONIC
o
DN65 - PN16,
DN80 - PN16/25,
DN100 - PN16 :
x4
7