Catalog Number - DeVilbiss Healthcare 525 Série Guide D'instructions

Concentrateur compact d'oxygène
Masquer les pouces Voir aussi pour 525 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
SPEcIFIcATIONS

catalog Number

Delivery Rate (Lower delivery rates available for
low flow applications)
maximum Recommended Flow(@ nominal outlet
pressures of zero and 7 kPa)
Outlet Pressure
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 21°C only)
0-1500 M (0-4921 ft)
1500-4000 M (4921-13123 ft)
Operating Environment Range*
5°C to 40°C, humidity range of 10% to 95%
Power Consumption
Weight
Sound Level (ISO 8359:1996 from front)
Dimensions
Pressure Relief Valve
Operating System
The visible "low oxygen" indicator will activate
at the following level
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
EmC Compliance To
*NOTE: The OSD performance at 5°C to 40°C, 95% R.H. through voltage range on the 525DS verified at 670m.
Specifications subject to change without notice.
EN - 8
DEvILBISS 5-LITER SERIES
525DS
1 to 5 LPM
5 LPM
8 .5 psig (58 .6 kPa)
115 V, 60 Hz, 3.3 Amp
97-127 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation of performance
No degradation in performance across the operating
voltage range .
310 Watts Average
275 Watts @ 2 .5 LPM & below
36 lbs . (16 .3 kilograms)
48 dbA
24 .5"H x 13 .5"W x 12"D (62 .2 x 34 .2 x 30 .4 cm)
40 psig±5psig (276 kPa±34 .5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
(The audible alarm will alert at approximately 75% . At less
than 60%, the red "service required" light will activate.)
-40°C to 70°C, humidity range of 10% to 100%, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part
CSA
CAN/CSA-C22 .2 No . 601 .1-M90
EN60601-1-2
525KS
1 to 5 LPM
5 LPM
8 .5 psig (58 .6 kPa)
220-230 V~, 50 Hz, 1.55 Amp
230 V~, 60 Hz, 1.9 Amp
187-255 V~, 50 Hz
195-255 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/50Hz only:
No degradation of performance
No degradation in performance across the operating
voltage range .
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @ 2 .5 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @ 2 .5 LPM & below
36 lbs . (16 .3 kilograms)
40 dbA (50Hz)
24 .5"H x 13 .5"W x 12"D (62 .2 x 34 .2 x 30 .4 cm)
40 psig±5psig (276 kPa±34 .5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
(The audible alarm will alert at approximately 75% . At less
than 60%, the red "service required" light will activate.)
-40°C to 70°C, humidity range of 10% to 100%, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part
TUV approved for 50 Hz only to
EN 60601+A1+A2
ISO8359: 1996
EN60601-1-2
SE-525

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

525ds525ks

Table des Matières