- Check the integrity of the device and its
accessories before and after each use in order
to detect any problem.
- If necessary, do not use the device and
replace any faulty component.
The P
2 is prepared for use as follows:
IEZOTOME
- Make sure that the device is connected
correctly and properly insulated.
- Switch the power switch (Fig. 2-5) to position I
(ON).
- Suspend the bottles or bags of physiological
saline or sterile water from the brackets.
- Raise the irrigation pump covers.
- Insert the irrigation line cassettes horizontally
into their housings.
- Close the covers.
- Insert the irrigation line perforating pins into
the bottles or pouches of physiological saline
and open the cap on the pins.
- Make any necessary setting adjustments
(irrigation flow rate, power mode, program,
etc.) (refer to chapter 7).
Note: This sequence must be completed under
conditions of controlled asepsis for the safety of
your patients. An operating assistant may be
needed.
Note: The previously-used display settings are
saved automatically when switching between
modes.
P
function
IEZOTOME
- Connect the P
IEZOTOME
of the connectors of the device (Fig. 1-7 and 8).
- Screw the selected tip onto the handpiece using
a
tip
wrench
recommendations of the clinical booklets).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the
P
2 handpiece.
IEZOTOME
2 handpiece cord to one
(according
to
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or
the multifunction footswitch (Fig. 8-2).
- When the irrigation solution reaches the end of
the tip, release the footswitch.
- Use the P
IEZOTOME
dental practice.
N
function
EWTRON
- Check that there are no traces of moisture on
the N
EWTRON
connections. Remove any moisture by wiping
and by drying with the multifunction syringe fed
with medical-quality filtered air.
- Connect the scaler cord to one of the
connectors of the device (Fig. 1-7 and 8).
- Align the electrical contacts of the N
handpiece and the scaler cord, then plug the
handpiece into the cord.
- Screw the selected tip onto the handpiece using
a tip wrench (according to the T
recommendations).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the
scaler cord.
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or
the multifunction footswitch (Fig. 8-2).
- When the irrigation solution reaches the end of
the tip, release the footswitch.
- Use the P
IEZOTOME
dental practice.
the
2 in accordance with good
handpiece or the scaler cord
2 in accordance with good
EWTRON