6. Commissioning, operation
Fitting of USB adapter cable
1. Switch the pressure transmitter to a depressurised state.
2. Remove the clamping nut and carefully pull out the instrument
EN
connector.
3. Connect the spade plug of the USB adapter cable to the instru-
ment's USB connection
4. Connect the USB connector of the adapter cable to the free
USB port of the computer.
5. Connect pressure transmitter to power supply.
6.5 Commissioning of USB instruments
For installing the driver, administrator rights are required.
Connect the USB connector with a USB 2.0 port to the computer.
Install the driver via the InstallWizard of the product software.
For further operation the P-3x data logger software is available (for details see instruction manual
P-3x data logger)
Details on the interface protocol or the DLL (Dynamic Link Library) are available on the software CD
and in the download section at www.wika.com.
All files and documents are available for download at www.wika.com.
6.6 Commissioning of CANopen instruments
Connect the CAN interface (e.g. PEAK PCAN-USB) to the PC and the pressure transmitter.
Start the CAN software (e.g. PEAK PCAN-View).
Make sure that the correct bit rate and node ID of the pressure transmitter is used.
For further information see "Quick Start Guide" under "Additional Instructions".