Errors and warning messages

Memory full
Reference memory full
You will also find an error number for the following messages in the Calibration/
Status menu to provide support for diagnostics:
Tolerance
Error 01
Please call Customer
Service (invalid)
Error 02
Error 03
Error 04
Error 05
Please observe the instructions on cleaning standards in the section on
Calibration.
56
Transfer the content of memory to a PC and then
delete the contents of memory.
A maximum of 50 references can be saved. It
may be necessary to delete old references.
Generally occurs only with major changes in
climatic or weather conditions. The deviation was
successfully compensated for by calibration and
correct measurements are still possible. You
should still recalibrate the device as soon as it is
operating in normal climatic conditions again,
however. If a change in climate cannot be
considered as the cause of the problem, you
should check whether the standard is clean.
Autodiagnosis has determined an impermissible
deviation in the measurement signal that cannot
be remedied by recalibrating.
Generally occurs when there is a significant
amount of dirt or dust on the standard or optics.
First try to clean the standard. You should only
have the optics cleaned by our Customer Service
department, for example as part of a yearly
recertification.
Defect in the electronics or operating error. First
check whether the standard is clean and whether
the device is properly snapped into the holder.
Defect in the lamp or electronics.
Defect in the electronics.