Electromagnetic Compatibility Information - Beurer medical BM 40 Mode D'emploi

Masquer les pouces Voir aussi pour BM 40:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24

Electromagnetic Compatibility Information

Table 1
Guidance and declaration of manufacturer – electromagnetic emissions
The device is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the device should assure that it
is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/fl icker emissions
IEC 61000-3-3
Table 2
Guidance and declaration of manufacturer – electromagnetic immunity
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Compliance
Electromagnetic environment – guidance
Group 1
The device uses RF energy only for its internal function. Therefore, its emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The device is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that sup-
Class A
plies buildings used for domestic purposes.
Compliance
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines ± 2 kV for power supply lines
± 1 kV line(s) to line(s)
<5 % U
(>95 % dip in U
)
T
T
for 0.5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
<5 % U
(>95 % dip in U
)
T
T
for 5 s
Compliance level
± 6 kV contact
± 8 kV air
± 1 kV line(s) to line(s)
<5 % U
(>95 % dip in U
)
T
T
for 0.5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
<5 % U
(>95 % dip in U
)
T
T
for 5 s
90
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
fl oors are covered with synthetic material, the rela-
tive humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the device requires continued operation during
power mains interruptions, it is recommended that
the device be powered from an uninterruptible
power supply or a battery.

Publicité

Table des Matières
loading

Table des Matières