Storage; Aging - Xylem YSI TruLine Série Mode D'emploi

Chaînes de mesure du ph à électrolyte polymère et gel
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TruLine
During short
measuring breaks
Overnight or
longer
T 814240
04/2014

Storage

Immerse the combination electrode in reference electrolyte (KCl 3 mol/L,
+
Ag
free). Prior to the next measurement, shortly rinse the combination
electrode with the test sample or deionized water.
Note
Prevent contact of the pH membrane to the beaker bottom to avoid scratches
on the pH membrane.
Put the clean combination electrode in the watering cap that is filled with
reference electrolyte (KCl 3 mol/L, Ag
Note
pH combination electrodes must not be stored dry or in deionized water. The
combination electrode could be permanently damaged by this. If the liquid in
the watering cap has dried up, condition the combination electrode in
reference electrolyte (KCl 3 mol/L, Ag
Note
During longer storing periods, salt sediments may develop on the watering
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the combination electrode is put into oper-
ation again.

Aging

Every pH combination electrode undergoes a natural aging process. With ag-
ing, the responding behavior becomes slower and the electrode slope and
asymmetry change. Moreover, extreme operating conditions can consider-
ably shorten the lifetime of the combination electrode. These are:
 Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
 High temperatures
 High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and
mechanical damage.
+
free).
+
free) for at least 24 hours.
Storage
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ysi truline 21Ysi truline 25Ysi truline 26Truline 27

Table des Matières