Xylem YSI TruLine Série Mode D'emploi page 6

Chaînes de mesure du ph à électrolyte polymère et gel
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Commissioning, measuring, calibration
Calibration
and measurement:
General rules
Subsequent
calibrations
6
be stored. Always keep the watering cap clean.
 TruLine 25, TruLine 26 and TruLine 21: Remove any gas bubbles behind
the pH membrane by shaking. With all other combination electrodes, gas
bubbles behind the pH membrane are not a problem.
 Connect the combination electrode to the meter.
 Calibrate the combination electrode according to the operating manual of
the meter and observe the following rules while doing so:
 Avoid the displacement of any solution (sample or buffer solution) from
one measurement to the next by taking the following measures:
– Shortly rinse the calibration and sample beakers with the solution the
beakers are to be filled with next.
– Between measurements, rinse the combination electrode with the so-
lution that follows. Alternatively, you can also rinse the combination
electrode with deionized water and then carefully dab it dry.
 To measure in aqueous solutions, it is recommended to immerse the com-
bination electrode in a vertical or slightly tilted position.
 Observe the correct depth of immersion and make sure the contact be-
tween the junction and test sample is thorough. The junction is in the area
of the bottom end of the shaft (see arrow).
TruLine 25
TruLine 26
Caution:
Only the shaft part of the combination electrode may be immersed!
 For measurements in aqueous solutions, provide approximately the same
stirring conditions for measuring as for calibrating.
The frequency of subsequent calibrations depends on the application. Many
meters provide an option where you can enter a calibration interval. After the
calibration interval has expired, the meter will automatically remind you of the
due calibration.
TruLine 21
TruLine 27
TruLine
T 814240
04/2014

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ysi truline 21Ysi truline 25Ysi truline 26Truline 27

Table des Matières