Error And Warning Messages; Maintenance; Technical Data - Velp Scientifica Arec 4 Manuel D'instructions

Table des Matières

4. Error and warning messages

When the control panel shows one of the following error messages, the instrument functions automatically stop.
Display indicator
Flashing Heating led and Hot plate led ON
Flashing Stirring led
To remove the error message, restart the instrument from the main switch.
If the alarm persists on the control panel, please contact VELP Scientifica's technical assistance service.

5. Maintenance

No routine or extraordinary maintenance is required except periodic cleaning.
Cleaning
Disconnect the unit from the power supply and use a cloth dampened with a non-flammable non-aggressive detergent.
Repair
Repairs must be carried out by authorized Velp personnel only.
The transport of the instrument by freight forwarders, couriers or others must be carried out using the original shockproof
packaging. Follow any instructions on the original packaging (e.g. palletizing).
It is the responsibility of the user to properly decontaminate the unit in case of hazardous substances remaining on the
surface or interior of the device. If in doubt about the compatibility of a cleaning or decontamination product, contact the
manufacturer or distributor of cleaning solution

6. Technical data

General features
Heating
Stir
Model
Voltage
Power input
Current consumption
Dimensions (WxHxD)
Weight
Construction material
Working in continuous
Maximum load on the plate
Noisiness
Environmental temperature admitted
Storage temperature admitted
Max humidity
Overvoltage category
Pollution degree CEI EN61010-1
Max altitude
Heating plate power output
Heating plate dimensions
Temperature range
Heat control type
Heating plate construction material
Overtemperature alarm
Stirring capacity
Speed range
Stir control type
Motor type
Motor rating output
Malfunction:
Heating plate overtemperature (T > 570 °C)
The stirring system doesn't run correctly
AREC 4
F20500670
Aluminum body – Technopolymer enclosure
7
230 V
50/60 Hz
280 W
1,2 A
160x100x270 mm (6,3x3,9x10,6 in)
1,9 kg (4,2 lb)
Admitted
25 kg
<< 80 dBa
+5...+40 °C
-10...+60 °C
80%
II
2
4000 m
250 W
100 x 100 mm (4"x4")
50 ÷ 550 °C
Analogic
Ceramic
Over 570°C
5 l H
O
2
100 ÷ 1500 rpm
Analogic
BLDC
10 W
F20510670
115 V
60 Hz
2,4 A
Table des Matières
loading

Ce manuel est également adapté pour:

Arec 7F20500670F20510670F20500680F20510680F20580680

Table des Matières